Features:

- Automatic and manual conductivity calibration up to 4 points
- Adjustable reference temperature and temperature coefficient
- Selectable cell constant and measurement unit
- Automatic / Manual temperature compensation and temperature calibration function
- Conductivity and total dissolved solids (TDS) conversion
- Four (4) TDS calibration curves for wide range of applications: EN27888, Linear, 442, NaCl
- Two (2) salinity calibration curves: NaCl, Seawater
- Built-in memory can store up to 500 data
- Auto-Hold / Auto-Stable measurement
- RS232 and USB outputs for connecting the meter to a printer or computer
- 2 years meter warranty

Specifications

:

Electrical conductivity

Measuring range
..uS/cm to 19.99 uS/cm; ..uS/cm to 1999.0 uS/ cm; ..uS/cm to 20.00 mS/cm; .. US/cm to 200.0 mS/cm
Resolution 0.05% over the measuring range (FS)
Accuracy +/- 0.6% across the measuring range (FS)
Reference temperature 5 to 30 C (adjustable)
Temperature Compensation 0.0 10.0% (adjustable)
Constant battery 0.1/1.0/10.0
Calibration points 4
Unit of measurement Automatically switches measuring range uS/cm or mS/cm or S/m
Total dissolved solids (TDS)
Measuring range
0:01 to 9.99 ppm, 0.1 to 999.9 ppm
1 ppm to 10:00 ppt; 10 ppm to 100.0 ppt
Resolution
0:01 ppm/0.1 ppt
Accuracy
+/- 0.1% across the measuring range (FS)
TDS Graphs

EN27888, 442, linear (0.40 ~ 1.0)
Resistivity
Measuring range
0.000 ohms / cm to 20,000 ohms/ cm; 0:00 ohms /cm to 200.0milliohm
/cm
Resolution
0.05% over the measuring range (FS)
Accuracy +/- 0.6% across the measuring range (FS)
Salinity
Measuring range
0.0 to 100.0 ppt; 0:00 to 10:00%
Resolution 0.1 ppt / 0.1%
Accuracy 0.2% across the measuring range (FS)
Calibration graph NaCl/sea
Temperature
Measuring range
-30.0 ~ 130 degrees C
Resolution
0.1 degrees C
Accuracy
+/- 0.4 C
Temperature Calibration Yes
Other features
Memory 500 data
Automatic Data logging Yes
Clock Yes
Auto-Hold Yes
Automatic shutdown Yes
Diagnostic messages Yes
Display LCD
Input BNC, phono, DC socket
Output RS232C, USB
Electrode stand Integrated
Power AC adapter 100-240V, 50/60Hz

Our Reviews



Recently Viewed

Instrument Choice will respond in 3 business hours

Get in touch with our expert scientists Instrument Choice.

In less than 3 business hours you will have an answer from our expert scientists.

Our Australian experts have access to a huge range of products from national and international suppliers.

If for any reason you have trouble using this form please email the following address and your enquiry will still receive a response from a scientist within 3 business hours: scientists@instrumentchoice.com.au