Temperature-RH Smart Sensor (8m cable) - S-THB-M008
Unfortunately, the S-THB-M008 has been discontinued and has been replaced by the S-THB-M008.


The 12-bit Temperature/RH Smart Sensor is designed to work with all Onset data loggers that accept Smart Sensors. All sensor parameters are stored inside the Smart Sensor, which automatically communicates configuration data information to the logger without any programming, calibration, or extensive user setup.
Specifications
- Measurement Range: Temp: -40?C to 75?C (-40?F to 167?F)RH: 0-100% RH at -40? to 75?C (-40? to 167?F); exposure to conditions below -20?C (-4?F) or above
95% RH may temporarily increase S-THB-M00X Figure 1the maximum RH sensor error by an additional 1%
- Accuracy: Temp: +/- 0.21?C from 0? to 50?C (0.38?F from 32? to 122?F) . See Figure 1.RH: +/- 2.5% from 10% to 90% RH (typical), to a maximum of +/-
3.5%. See Figure 2 for full range.
- Resolution:Temp: 0.02?C at 25?C (0.04?F at 77?F) See Figure 1RH: 0.1% RH at 25?C (77?F)
- S-THB-M00X Figure 2Bits Per Sample:Temp: 12RH: 10
- Drift:Temp: < 0.1?C (0.18?F) per yearRH: < 1% per year typical; hysteresis 1%
- Response Time:Temp: 5 minutes in air moving 1 m/secRH: 5 minutes in air moving 1 m/sec with protective cap
- Operating Temperature Range: -40?C to 75?C (-40?F to 167?F)
- Environmental Rating Weatherproof: 0 to 100% RH intermittent condensing environments. For best results, the Temp/RH Smart Sensor should be mounted
inside a protective enclosure, such as a solar radiation shield.
- Housing PVC cable jacket with ASA styrene polymer RH sensor cap; modified hydrophobic polyethersulfone membrane
- Sensor Dimensions: 10 x 35 mm (0.39 x 1.39 in)
- Weight: - 110 g (3.88 oz)
- Number of Data Channels: 2
- Measurement Averaging Option: No
- Cable Lengths Available: 2 Meter and 8 Meter
- Length of Smart Sensor Network Cable: 0.5 m (1.6 ft)
This product meets CE specification EN61326 criterion C for ESD, criterion C for Radiated Immunity, criterion C for Fast Transient, criterion B for Conducted Immunity, criterion A for Power Frequency Magnetic Fields, and Radiated Emissions Group 1, Class B. To minimize measurement errors due to ambient RF, use the shortest possible probe cable length and keep the probe cable as far as possible from other cables.
Note: Sensor requires protection from rain or direct splashing; Radiation shield (RS3) strongly recommended for use in sunlight; sensors can be used in intermittent condensing environments up to 30?C and non-condensing above 30?C.
Our Reviews
Recently Viewed
Instrument Choice will respond in 3 business hours
Get in touch with our expert scientists Instrument Choice.
In less than 3 business hours you will have an answer from our expert scientists.
Our Australian experts have access to a huge range of products from national and international suppliers.
If for any reason you have trouble using this form please email the following address and your enquiry will still receive a response from a scientist within 3 business hours: scientists@instrumentchoice.com.au