Aquaculture Photometer - IC-HI83303
Product code: IC-HI83303
3+$989.303ex. GST
5+$974.0045ex. GST
10+$963.8055ex. GST
The HI83303 benchtop photometer measures 12 different key water quality parameters using 20 different methods. This photometer features an innovative optical system that uses LEDs, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.
Made with the aquaculture industry in mind, the HI83303 is a comprehensive solution to maintaining optimal chemical and environmental conditions, preventing disease and increasing production. The HI83303 measures vital parameters such as alkalinity, calcium, nitrite, and phosphate. Alkalinity plays a part in a dynamic relationship with pH and CO2 concentrations, high alkalinity water lowers fluctuations in pH. The buffering capacity acts to store extra CO2 essential for photosynthesis in the ponds to produce oxygen. Maintaining calcium at certain levels is vital to proper fish growth and development. Excessive nitrite can be toxic to fish. When nitrite interacts with hemoglobin the iron becomes oxidized and the blood cell can no longer carry oxygen. Phosphate is essential to plant growth; too much phosphate in an aquaculture system can contribute to algal blooms decreasing dissolved oxygen vital for a successful ecosystem.
A digital pH electrode input allows the user to measure pH by a traditional glass electrode. The digital pH electrode has a built-in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built-in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.
The HI83303 offers an absorbance measuring mode that allows for CAL Check standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the five wavelengths of light (420 nm, 466 nm, 525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for educators to teach the concept of absorbance by using the Beer-Lambert Law.
Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.
Features:
- Backlit graphic display allows for easy viewing in low light conditions
- The 128 x 64 Pixel LCD allows for a simplified user interface with virtual keys and on-screen help to guide the user through use
of the meter
- The measurement is taken after the countdown timer expires.
- Countdown timer ensures that all readings are taken at the appropriate reaction intervals regardless of user for better consistency
in measurements
- Hanna's exclusive CAL Check cuvettes for validation of light source and detector
- Allows for the user to plot concentration versus absorbance for a specific wavelength for use with user-supplied chemistry or for
teaching principles of photometry
- Appropriate unit of measure along with chemical form is displayed along with reading
- Automatically convert readings to other chemical forms with the touch of a button
- Aids in preventing stray light from affecting measurements
- Logged readings can be quickly and easily transferred to a flash drive using the USB-A host port or to a computer using the micro
USB-B port
- Data is exported as a .CSV file for use with common spreadsheet programs
Digital pH Electrode Input:
- Measure pH and temperature with a single probe
- Good Laboratory Practice (GLP) to track calibration information including date, time, buffers used, offset and slope for traceability
- pH CAL Check alerts user to potential problems during the calibration process
- Space saving having a pH meter and photometer built into one meter
Data Logging:
- Up to 1000 photometric and pH readings can be stored by simply pressing the dedicated LOG button. Logged readings are just as easily
recalled by pressing the RCL button
- Sample ID and User ID information can be added to a logged reading using alphanumeric keypad
Specifications
pH:
- pH Range: Photometer: 6.5 to 8.5 pH
- pH electrode: -2.00 to 16.00 pH
- pH Resolution: Photometer: 0.1 pH
- pH electrode: 0.01 pH
- pH Accuracy: Photometer: 0.1 pH
- pH electrode: +/- 0.01 pH
- pH Calibration: Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01)
- pH Temperature Compensation: Automatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used
- pH CAL Check (electrode diagnostics): clean electrode and check buffer/check probe displayed during calibration
- pH Method: Photometer: phenol red
- pH-mV Range: +/- 1000 mV
- pH-mV Resolution: 0.1 mV
- pH-mV Accuracy: +/- 0.2 mV
Oxygen, Dissolved:
- Oxygen, Dissolved Range: 0.0 to 10.0 mg/L (as O2)
- Oxygen, Dissolved Resolution: 0.1 mg/L
- Oxygen, Dissolved Accuracy: +/- 0.4 mg/L, +/-3% of reading
- Oxygen, Dissolved Method: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified
Winkler method
Absorbance:
- Absorbance Range: 0.000 to 4.000 Abs
- Absorbance Resolution: 0.001 Abs
- Absorbance Accuracy: +/-0.003Abs at 1.000 Abs
Alkalinity:
- Alkalinity Range: Freshwater: 0 to 500 mg/L (as CaCO3)
- Seawater: 0 to 300 mg/L (as CaCO3)
- Alkalinity Resolution: 1 mg/L
- Alkalinity Accuracy: +/- 5 mg/L, +/- 5% of reading
- Alkalinity Method: Colorimetric method
Ammonia:
- Ammonia Range (Low Range): 0.00 to 3.00 mg/L
- Ammonia Range (Medium Range): 0.00 to 10.00 mg/L
- Ammonia Range (High Range): 0.0 to 100.0 mg/L
- Ammonia Resolution: Low and Medium Range: 0.01 mg/L
- High Range: 0.1 mg/L
- Ammonia Accuracy (Low Range): +/0.04 mg/L, +/- 4% of reading
- Ammonia Accuracy (Medium Range): +/- 0.05 mg/L, +/- 5% of reading
- Ammonia Accuracy (High range): +/- 0.5 mg/L, +/- 5% of reading
- Ammonia Method: Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method
Calcium:
- Calcium Range (Freshwater): 0 to 400 mg/L (as Ca2+)
- Calcium Range (Seawater): 200 to 600 mg/L (as Ca2+)
- Calcium Resolution: 1 mg/L
- Calcium Accuracy (Freshwater): +/- 10 mg/L, +/- 5% of reading
- Calcium Accuracy (Seawater): +/- 6% of reading
- Calcium Method: Freshwater: Adaptation of the Oxalate method
Chlorine:
- Free Chlorine Range: 0.00 to 5.00 mg/L (as Cl2)
- Free Chlorine Resolution: 0.01 mg/L
- Free Chlorine Accuracy: +/- 0.03 mg/L, +/- 3% of reading
- Total Chlorine Range: 0.00 to 5.00 mg/L (as Cl2)
- Total Chlorine Resolution: 0.01 mg/L
- Total Chlorine Accuracy: +/- 0.03 mg/L, +/- 3% of reading
- Chlorine Method: Adaptation of the EPA 330.5 DPD method
Copper:
- Copper Range (Low Range): 0.000 to 1.500 mg/L (as Cu2+)
- Copper Range (High range): 0.00 to 5.00 mg/L (as Cu2+)
- Copper Resolution: 0.001 mg/L; 0.01 mg/L
- Copper Accuracy (Low Range): +/- 0.01 mg/L, +/- 5% of reading
- Copper Accuracy (High Range): +/- 0.02 mg/L, +/- 4% of reading
- Copper Method: Adaptation of the EPA bicinchoninate method
Nitrate:
- Nitrate Range: 0.0 to 30.0 mg/L (as NO3-- N)
- Nitrate Resolution: 0.1 mg/L
- Nitrate Accuracy: +/- 0.5 mg/L, +/- 10% of reading
- Nitrate Method: Adaptation of the cadmium reduction method
Nitrite:
- Nitrite Range (Freshwater, Low Range): 0 to 600 ug/L (as NO2--N)
- Nitrite Range (Freshwater, High Range: 0 to 150 mg/L (as NO2-)
- Nitrite Range (Seawater Ultra Low Range): 0 to 200 ug/L (as NO2--N)
- Nitrite Resolution (Freshwater): 1 ug/L; 1 mg/L
- Nitrite Resolution (Seawater): 1 ug/L
- Nitrite Accuracy (Freshwater, Low Range): +/- 20 ug/L, +/- 4% of reading
- Nitrite Accuracy (Freshwater, High Range): +/- 4 mg/L, +/-4% of reading
- Nitrite Accuracy (Seawater): +/- 10 ug/L, +/- 4% of reading
- Nitrite Method: Low Range and Seawater: Adaptation of the EPA Diazotization method 354.1
- High Range: Adaptation of the Ferrous Sulfate method
Phosphate:
- Phosphate Range (Freshwater, Low Range): 0.00 to 2.50 mg/L (as PO43-)
- Phosphate Range (Freshwater, High range: 0.0 to 30.0 mg/L(as PO43-)
- Phosphate Range (Freshwater, Seawater): Ultra Low Range: 0 to 200 g/L (as P)
- Phosphate Resolution (Freshwater): 0.01 mg/L; 0.1 mg/L
- Phosphate Resolution (Seawater): 1 ug/L
- Phosphate Accuracy (Freshwater, Low Range): +/- 0.04 mg/L, +/- 4% of reading
- Phosphate Accuracy (Freshwater, High Range): +/- 1 mg/L, +/- 4% of reading
- Phosphate Accuracy (Seawater, Ultra Low Range): +/- 5 ug/L, +/- 5% of reading
- Phosphate Method: Freshwater Low Range: Adaptation of the Ascorbic Acid method
- Freshwater High Range and Seawater Ultra Low Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater,
18th edition, Amino Acid method
- Seawater: Adaptation of the Zincon method
File | Type | Size | Created At |
---|---|---|---|
man8330.._19.pdf | 6.34 MB | 27/7/2022, 2:27 PM |
Delivery Details
FREE Startrack Premium Air Express in Australia - Non-Bulky Items |
---|
Free Delivery in Australia for orders over $100. Overnight delivery to most capital cities and surrounding areas (2 to 5 day delivery outside these locations). - Most articles delivered next business day when using this service. Next–business-day delivery is only available between capital cities (excluding Darwin, and for Perth between CBDs only) and some major centres. - Shipped from Instrument Choice on South Australian business days
Free Delivery in Australia for orders over $100
|
For alternate delivery options and international shipping, click here.
FREE Express Road Courier (NO PO Box, Authority to leave) - Bulky Items |
---|
Free Delivery in Australia for orders over $100. Typically 2 to 5 working day delivery time to capital cities (Up to 2 weeks outside these locations.)
- Products are shown as bulky items on their product pages - Delivered only on business days at your location, delivery unavailable on weekends and public holidays
- Shipped from Instrument Choice on South Australian working days
This service does not deliver to Post Office Boxes
Free Delivery in Australia for orders over $100
|
For alternate delivery options and international shipping, click here.
Why Instrument Choice?
#1. SUPER-RESPONSIVE CUSTOMER SERVICE
#2. SPECIAL ADVICE FROM SCIENTISTS
#3. Easy Shopping
#4. Free Express Shipping
Why pay for shipping when Instrument Choice has negotiated such amazing freight rates for Australia and New Zealand? Overnight shipping is available to capital cities for 93% of stocked items, the rest still travel by express road to get to you quickly.
Get details about our super-fast and secure shipping options here