Trouble with pH meter calibration
Trouble with pH meter calibration |
The Problem: | |
Nick called and was having trouble calibrating his pH meter. He had pH4 and pH7 calibration solutions but was unable to get the pH meter to calibrate. He was worried he may have damaged the probe and thought this may be why it wasn’t calibrating, but he could not remember knocking the probe at all. | |
The Solution: | |
We discussed Nick’s calibration procedure and it turned out he was reusing the same buffer solution each time he calibrated. After each calibration he was tipping the solution back into the same bottle. This repeated recycling contaminated the pH4 and pH7 solutions meaning that they no longer had pH values of 4 and 7. We recommended that Nick replace his pH4 and pH7 buffer solutions and the issue resolved itself. Below are some more tips about calibrating your pH probe.
If you need any assistance with pH meter calibration or if you wish to know further details about calibration please feel free to contact one of our friendly Scientists via email or phone on 1300 737 871. | |