The design of the EC-SaltTestr11 Waterproof salinity meter enables quick and accurate salinity measurements. This user-friendly meterguarantees years of trouble-free operation and requires very little maintenance. One critical step to keep this meter performing to the highest level of accuracy is to perform regular calibration. Thankfully, calibration of the EC-SaltTestr11 is an incredibly simple task that takes only eight steps.

 

Figure 1 Watch an Instrument Choice unbox and demonstrate how to calibrate the EC-SaltTestr11 Salinity Meter here.

How to Calibrate the EC-SaltTestr11 Waterproof Salt Meter

You will need:

Method

  1. Half fill both the containers with the NaCI calibration solution. One container is for pre-rinsing, and the other is to perform the calibration.
  2. Start with the meter switched on and in measurement mode.
  3. Remove the cap on the top of the device. Under the cap is the two buttons required for calibration, these are increase and decrease selection.
  4. Remove the end cap covering the electrode and then rinse the electrode in the first container of the calibration standard.
  5. Move the electrode into the second container of the calibration solution and swirl gently to create a homogenous sample. Allow some time for the reading to stabilise.
  6. Press the increase or decrease button to enter the calibration mode. 'CAL' will appear on the top of the LCD. When in calibration mode, the upper display shows the measured reading of the solution based on previous calibrations (if any), and the lower display shows the default uncalibrated reading.
  7. Use the increase and decrease buttons to adjust the upper display to the correct value of the calibration solution.
  8. After adjusting to match the calibration solution value, press no buttons and wait five seconds for the tester to confirm the calibration by displaying 'CO' automatically.
  9. The salinity meter will then automatically return to measurement mode, and the calibration is complete. Replace the top lid, and the meter is ready to take accurate salinity measurements.

Conclusion

The EC-SaltTestr11 is a straightforward salinity meter that can simply be user calibrated to ensure consistently accurate measurements. If you would like more information on the EC-SaltTestr11 salinity meter or want to know more about salinity calibration solutions, see the additional links (below) or contact us!

Additional links

Also interesting

How to Calibrate Sound Level on the C-DSM1 Sound Level Meter

Calibration is an essential task to ensure your sound level meter is collecting accurate and repeatable results. Performing calibration for the C-DSM1 takes less than five minutes, which is perfect if your time is at a premium!

Read More
How to Calibrate Sound Level on the C-DSM1 Sound Level Meter

How to Assemble the IC0370 Wireless Weather Station

The IC0370 wireless weather station is a feature-rich and highly functional instrument that’s guaranteed to impress weather enthusiasts with its professional standard reliability and accuracy. It boasts an impressive range of features and is incredibly easy to set up and get started.

Read More
How to Assemble the IC0370 Wireless Weather Station